ICU Ventilator S1600

Application
The ventilator is suitable for various kinds of medical institutions for cardiopulmonary resuscitation respiratory support. Acute respiratory failure caused by various reasons or incomplete oxygenation dysfunction. Intra-operation,  post-operation respiratory support, other respiratory treatment.


Trust Points

  • Simplicity: easy to use, easy to move with 4 wheels.
  • Choice: adapt the equipment to your patients and procedures freely
  • Patient Centered Ventilation: Precision in an anesthesia ventilator, from conventional ventilation to advanced modes, including 5 modes: IPPV; A/C; SIMV; SPONT/CPAP; MANU.
  • Designed and manufactured by Perlong Medical with over 23 years experience in this area.
  • Flexible configurations to suit your needs.
  • International standard and advanced technology suitable for wide range use.
  • Compact interface and big screen give you better operating experience.
  • Over 2,000 units installed in the world.

Features

  • 10.4” TFT LCD screen displays the Ventilation parameters, Alarming information and Waveform.
  • Recycle breathing pipe, ensure easy operating and keep tidy.
  • Multiple working modes such as volume control and pressure limit, adapt to wide range patient.
  • Vaporizer with temperature, flow compensation and self-lock function, keep safety anytime.
  • Multiple parameters monitoring interface, make every parameter clear, let users know the patient conditions in all aspects;
  • Real time pressure-time, flow-time loop graphics and high precision O2 concentration detection function included.
  • Venturi effect technology ensure the machine working without air gas source.

Safety

  • Three level alarming system, visual and sound alarm information.
  • With lots of alarming, reminding and protection functions.
  • Advanced power management control technology.
  • With built-in backup power source, when outside power source goes off, back-up power source starts to work.
  • Self-check before running, eliminate system mistake.
  • Separate design of electric and gas, keep safety running of ventilator.

Specifications


Ventilation modes:

IPPV; A/C; SIMV; SPONT/CPAP; MANU

Ventilator parameter ranges

 

Tidal volume(Vt)

0, 20 mL 1500 mL

FrequenceFreq

1bpm 100 bpm

I/E

4:11:8

PEEP

0cmH2O 20 cmH2O

Pressure triggering sensitivityPTr

20 cmH2O 0 cmH2OBased on PEEP

SIGH

0(off)  1/100 5/100

Apnea Ventilation

OFF5 s ~ 60 s

Monitored Parameters

 

FrequenceFreq

0 /min 100 /min

Tidal volume(Vt)

0 mL 2000 mL

MV
Airway pressure

0 L/min - 99 L/min
0cmH2O 100cmH2O

Oxygen concentration

15 % 100 %

Graphical display:

 

P-T(pressure – time)

 

F-T(flow – time)

 

 

Size
1.Wooden case packing size(main engine)L 560*W 560*H 615mm , GW35KG ; NW17KG
2.Wooden case packing size(trolley)L650*W690*H 520 mm , GW38KG NW18G

Alarm and protection 

The AC power failure alarm

Power failure or no connection

Internal battery backup low voltage alarm

11.3±0.3V

No tidal volume

No tidal volume within 6s

High oxygen concentration alarm
Low oxygen concentration alarm

19%-100%
18%-99%

High Airway  pressure alarm
Low Airway  pressure alarm
High Minute Volume alarm
Low Minute Volume alarm
Continuous Pressure alarm

20cmH2O-100cmH2O
0cmH2O-20cmH2O
1L/min-99L/min
0-20L/min
(PEEP+1.5kPa) over 16s

Suffocation warning

5s-60s no spontaneous ventilation

 

Working conditions

 

Gas source

O2,

Pressure

280kPa-600kPa

Voltage

-220V±22V

Power frequency

50Hz±1Hz

Input power

40VA